Kelvin-S100
#
DescriptionAdapt’s flagship product designed and built to escort critically sensitive pharmaceutical products. A true end-to-end cold chain solution for monitoring live assets travelling by road or air. Include Kelvin while packing, and monitor your air cargo consignments with Zero Human Intervention.
#
FeaturesTemperature Monitoring
Real time location tracking
Waterproof and Food grade packing
Flight mode
LCD Display for Temperature Reading
Reliable and cost effective
Geofencing
High-Precision Digital Sensor
IP64 Grade
#
Product Anatomy#
Technical SpecificationsRecording Options | Single-Use |
Temperature Range | –20 °C to +60 °C |
Temperature Accuracy | ±0.5°C (from –20 °C to +60 °C) |
Temperature Resolution | 0.1 °C |
Data Storage Capacity | 31,200 readings |
Shelf Life / Battery | 2 years / CR2450 button cell |
Reading Interval | 10 minutes (Minimum Configurable 5 min) |
Reading Trasmission | 30 minutes (minimum configurable) |
Startup Mode | Button Press for 3 seconds. |
Stop Mode | Button Press for 3 seconds, or Plug the USB to a Port |
Water Resistance | IP64 |
Weight | 111 g |
Dimensions | 101mm x 50mm x 18.8mm (L × W × H) |
Battery | Li-on, 3000 mAh |
Runtime(Max Trip duration) | 30days @ 60 Min(minimum) sending Interval at room temperature & collection Interval 10 min. |
WIFI | 2.4 Ghz |
Cellular | 2G/4G(Cat M1), Sim Inbuilt |
Temperature Calibration | Factory Calibrated |
Temperature Sensor Type | Digital |
Probe Support | No |
Standards | FCC/CE Compliance |
Charging Time | upto 8 hours for full charge |
Charging voltage | 5v DC, USB type C |
Violations | Violation are logged as events on web portal and devices |
Notifications | Alerts as SMS, Email,Pop-up(Portal side) |
Compliance | FAA:AC 91.21 -1D, 2017, p. 91 EU: AMC1 CAT.GEN.MPA.140 As per the guidelines of IATA on Portable Electronics Devices(PEDs)Airline Approved |
Report File Output | Graphs, PDFs, CSVs (downloadable from portal side) |